A platform for research: civil engineering, architecture and urbanism
Hand-held metal detectors and medical devices: measurements and testing for electromagnetic compatibility
This work examines the electromagnetic compatibility (EMC) of several priority medical devices, such as implanted cardiac pacemakers and implanted nerve stimulators, with the emissions from 28 different hand-held metal detectors (HHMDs). The HHMD emissions were measured and mapped to assess the waveforms, magnitude, and distribution of emission field strengths. Testing with the sample medical devices was performed using a saline filled torso simulator for the implantable type devices. Emissions from the HHMDs were observed to disrupt the function of some of the sample medical devices. One HHMD exhibited significantly higher emissions than the other metal detectors, and caused significant disruptions on several of the sample medical device. The findings and data from this work are being used to help develop standards for characterizing the performance of medical devices.
Hand-held metal detectors and medical devices: measurements and testing for electromagnetic compatibility
This work examines the electromagnetic compatibility (EMC) of several priority medical devices, such as implanted cardiac pacemakers and implanted nerve stimulators, with the emissions from 28 different hand-held metal detectors (HHMDs). The HHMD emissions were measured and mapped to assess the waveforms, magnitude, and distribution of emission field strengths. Testing with the sample medical devices was performed using a saline filled torso simulator for the implantable type devices. Emissions from the HHMDs were observed to disrupt the function of some of the sample medical devices. One HHMD exhibited significantly higher emissions than the other metal detectors, and caused significant disruptions on several of the sample medical device. The findings and data from this work are being used to help develop standards for characterizing the performance of medical devices.
Hand-held metal detectors and medical devices: measurements and testing for electromagnetic compatibility
Witters, D.M. (author) / Buzduga, V. (author) / Seidman, S. (author) / Kainz, W. (author) / Casamento, J. (author) / Ruggera, P. (author)
2005-01-01
1934722 byte
Conference paper
Electronic Resource
English
British Library Conference Proceedings | 2005
|Pre-prototype hand-held fire detectors
Engineering Index Backfile | 1966
|Electromagnetic compatibility testing of vehicles gets more complex
British Library Online Contents | 2009
Electromagnetic Compatibility (EMC)
British Library Online Contents | 1997
|Electromagnetic Compatibility (EMC)
British Library Online Contents | 1995