Pulse oximeter bench tests under different simulated skin tones
AbstractPulse oximeters’ (POs) varying performance based on skin tones has been highly publicised. Compared to arterial blood gas analysis, POs tend to overestimate oxygen saturation (SpO2) values for people with darker skin (occult hypoxemia). The objective is to develop a test bench for assessing commercial home and hospital-based POs in controlled laboratory conditions. A laboratory simulator was used to mimic different SpO2 values (~ 70 to 100%). Different neutral density and synthetic melanin filters were used to reproduce low signal and varying melanin attenuation levels. Six devices consisting of commercial home (Biolight, N = 13; ChoiceMMed, N = 18; MedLinket, N = 9) and hospital-based (Masimo Radical 7 with Neo L, N = 1; GE B450 Masimo SET with LNCS Neo L, N = 1; Nonin 9550 Onyx II™, N = 1) POs were reviewed and their response documented. Significant variations were observed in the recorded SpO2 values among different POs when exposed to identical simulated signals. Differences were greatest for lower SpO2 (< 80%) where empirical data is limited. All PO responses under low signal and melanin attenuation did not change across various simulated SpO2 values. The bench tests do not provide conclusive evidence that melanin does not affect in vivo SpO2 measurements. Research in the areas of instrument calibration, theory and design needs to be further developed. Graphical Abstract
Pulse oximeter bench tests under different simulated skin tones
AbstractPulse oximeters’ (POs) varying performance based on skin tones has been highly publicised. Compared to arterial blood gas analysis, POs tend to overestimate oxygen saturation (SpO2) values for people with darker skin (occult hypoxemia). The objective is to develop a test bench for assessing commercial home and hospital-based POs in controlled laboratory conditions. A laboratory simulator was used to mimic different SpO2 values (~ 70 to 100%). Different neutral density and synthetic melanin filters were used to reproduce low signal and varying melanin attenuation levels. Six devices consisting of commercial home (Biolight, N = 13; ChoiceMMed, N = 18; MedLinket, N = 9) and hospital-based (Masimo Radical 7 with Neo L, N = 1; GE B450 Masimo SET with LNCS Neo L, N = 1; Nonin 9550 Onyx II™, N = 1) POs were reviewed and their response documented. Significant variations were observed in the recorded SpO2 values among different POs when exposed to identical simulated signals. Differences were greatest for lower SpO2 (< 80%) where empirical data is limited. All PO responses under low signal and melanin attenuation did not change across various simulated SpO2 values. The bench tests do not provide conclusive evidence that melanin does not affect in vivo SpO2 measurements. Research in the areas of instrument calibration, theory and design needs to be further developed. Graphical Abstract
Pulse oximeter bench tests under different simulated skin tones
Med Biol Eng Comput
Swamy, Suvvi K. Narayana (author) / He, Chenyang (author) / Hayes-Gill, Barrie R. (author) / Clark, Daniel J. (author) / Green, Sarah (author) / Morgan, Stephen P. (author)
2024-04-24
Article (Journal)
Electronic Resource
English
The Light Source Color Preferences of People of Different Skin Tones
British Library Online Contents | 2002
|Design of Wireless Intra-Operative Pulse Oximeter With Reticulated Pressure-Sensitive Head
British Library Online Contents | 2015
|Comparison of simulated wet bench fires with small- and intermediate-scale fire tests
Online Contents | 2008
|Comparison of simulated wet bench fires with small- and intermediate-scale fire tests
British Library Online Contents | 2008
|