Comparison of Two Ambient Beta Gauge PM10 Samplers
This paper reports on the performance of the Kimoto 180 sampler and the Wedding ambient PM10 beta gauge sampler. Monodisperse ammonium fluorescein test particles were generated in the laboratory and used to determine the penetration curve of the Kimoto 180 cyclonic inlet. It was found that the actual DpaSOof the Kimoto 180 inlet, 3.5 (xm, is much lower than the designated standard value, 10 um. In the field test, the two beta gauge samplers were collocated with an Andersen SA1200 high-volume sampler to compare their measured daily average PM10 concentrations.
The low Dpa50 of the Kimoto 180 inlet serves to explain why its daily average PM10 concentrations were much lower than the actual PM10 concentrations found in the field study. In addition, the PM10 concentrations of the Kimoto 180 beta gauge sampler were found to be seriously affected by the water vapor content of the ambient air. In contrast, the daily average PM10 concentrations of the Wedding beta gauge sampler were found to be more accurate, and influences by ambient conditions were insignificant
Comparison of Two Ambient Beta Gauge PM10 Samplers
This paper reports on the performance of the Kimoto 180 sampler and the Wedding ambient PM10 beta gauge sampler. Monodisperse ammonium fluorescein test particles were generated in the laboratory and used to determine the penetration curve of the Kimoto 180 cyclonic inlet. It was found that the actual DpaSOof the Kimoto 180 inlet, 3.5 (xm, is much lower than the designated standard value, 10 um. In the field test, the two beta gauge samplers were collocated with an Andersen SA1200 high-volume sampler to compare their measured daily average PM10 concentrations.
The low Dpa50 of the Kimoto 180 inlet serves to explain why its daily average PM10 concentrations were much lower than the actual PM10 concentrations found in the field study. In addition, the PM10 concentrations of the Kimoto 180 beta gauge sampler were found to be seriously affected by the water vapor content of the ambient air. In contrast, the daily average PM10 concentrations of the Wedding beta gauge sampler were found to be more accurate, and influences by ambient conditions were insignificant
Comparison of Two Ambient Beta Gauge PM10 Samplers
JinnTsai, Chuen (author) / Cheng, Yu-Hsiang (author)
Journal of the Air & Waste Management Association ; 46 ; 142-147
1996-02-01
6 pages
Article (Journal)
Electronic Resource
Unknown
Performance Characteristics of PM10 Samplers under Calm Air Conditions
Taylor & Francis Verlag | 2000
|Use of the Aerodynamic Particle Sizer to Measure Ambient PM10–2.5: The Coarse Fraction of PM10
Taylor & Francis Verlag | 2006
|Taylor & Francis Verlag | 2000
|Ambient PM10 and respiratory illnesses in Colombo City, Sri Lanka
Online Contents | 2008
|Use of passive ambient ozone (O3) samplers in vegetation effects assessment
Online Contents | 2001
|